HP 1GSa/s Digitizing Oscilloscope 16532A manual A/D Test Run Screen

Models: 1GSa/s Digitizing Oscilloscope 16532A

1 90
Download 90 pages 21.04 Kb
Page 73
Image 73

HP 16532A - Service

7. Touch the A/D Test field.

The individual test run screens, see figure 6-7, give the test name, a brief description of the test, number of test runs, and the number of test failures.

Figure 6-7. A/D Test Run Screen

8. Touch Run, then drag your finger to Single or Repetitive.

During the time a Single run or a Repetitive run is executing, the Run field displays Stop. 9. To stop a Repetitive run, touch Stop. See figure 6-8.

Figure 6-8. Stop Field

6-7

Page 73
Image 73
HP 1GSa/s Digitizing Oscilloscope 16532A manual A/D Test Run Screen