Getting Started
22 AFG3000 Series Quick Start User Manual
The following table shows the output window (maximum and minimum levels) for sine waveform when you change the load
impedance (L). Load impedance will affect the output window. Refer to page 54 for setting up load impedance.

Overheat Protection (AFG3011 Only)

The instrument internal temperature is monitored in the AFG3011. A warning message will appear if the internal tempera-
ture reaches a threshold level, and signal output will automatically turn off. If the warning message appears, check for the
following conditions (see page 6 for operating requirements):
The ambient temperature requirement is being met.
The required cooling clearance is being met.
The instrument fan is working properly.
L = 50 ΩL = High Z
AFG3011
Maximum level
Minimum level
(Maximum amplitude)
10 V
–10 V
(20 Vp-p)
20 V
–20 V
(40 Vp-p)
AFG3021B/AFG3022B
Maximum level
Minimum level
(Maximum amplitude)
5 V
–5 V
(10 Vp-p)
10 V
–10 V
(20 Vp-p)
AFG3101/AFG3102
Maximum level
Minimum level
(Maximum amplitude)
10 V
–10 V
(10 Vp-p)
10 V
–10 V
(20 Vp-p)
AFG3251/AFG3252
Maximum level
Minimum level
(Maximum amplitude)
5 V
–5 V
(5 Vp-p)
10 V
–10 V
(10 Vp-p)