Specifications
AFG3000 Series Quick Start User Manual 85
Amplitude (50 Ω)
Range420 mVp-p to 20 Vp-p
Accuracy5
(Amplitude: 10 Vp-p)
±(2% of setting + 2 mV)
(1 kHz sine waveform, 0 V offset, >20 mVp-p amplitude)
Resolution 0.1 mVp-p, 0.1 mVrms, 1 mV, 0.1 dBm or 4 digits
Units6Vp-p, Vrms, dBm, and Volt (High Level and Low Level)
Output Impedance 50 Ω
Isolation 42 Vpk maximum to earth
DC Offset (50 Ω)
Range7±10 Vpk ac + dc into 50 Ω
Accuracy8
(|setting| 5V)
±(2% of |setting| +10 mV + 1% of amplitude (Vp-p))
Resolution 1 mV
Internal Noise Add
Range 0.0% to 50% of amplitude setting (Vp-p) of signal waveform
Resolution 1%
1. Triggered/Gated Burst Mode: 1 μHz to 5 MHz
2. Triggered/Gated Burst Mode: 1 mHz to 2.5 MHz
3. Resolution: 0.01° (sine), 0.1° (other standard waveforms)
4. 40 mVp-p to 40 Vp-p into open circuit load
5. ±(2% of setting + 2 mV) (Typical) (Amplitude: >10 Vp-p)
6. dBm is used only for sine waveform, and Vrms is not available for Arb and Noise waveforms.
7. ±20 Vpk ac + dc into open circuit load
8. ±(2% of |setting| + 10 mV + 1% of amplitude (Vp-p)) (Typical) (|setting| >5 V)
Add 1.0 mV per °C for operation outside the range of 20 °C to 30 °C.
Waveforms (cont.)