Getting Started
AFG3000 Series Quick Start User Manual 21

Equivalent Output Circuits

The following illustrations show the equivalent output circuits for the AFG3000 series instruments:
1. AFG3011
Output signals do not exceed ±20 V
when the >50 Ω load impedance is
used.
2. AFG3021B and AFG3022B
Amplitude and offset of the output
signals are not affected by load
impedance.
3. AFG3101/AFG3102
Output signals do not exceed ±10 V
when the >50 Ω load impedance is
used.
Voltage over the maximum level is
clipped.
Amplitude and offset are affected
when you change the load imped-
ance. The maximum and minimum
levels do not exceed ±10 V, respec-
tively.
4. AFG3251/AFG3252
Output signals do not exceed ±10 V
when the >50 Ω load impedance is
used.
Offset
50 Ω
Signal
Offset
50 Ω
Signal
Offset
50 Ω
Signal
Offset
Signal
50 Ω
Current source Voltage source