Operating Basics
AFG3000 Series Quick Start User Manual 61
Quick Tip
To output continuous waveforms by synchronous operation, select External as the trigger source for the master unit to
stop signal generation. Select Inf-Cycles as the burst count for both master and slave units. Change the trigger source
of the master unit to Internal to restart the signal generation.
7. Select the trigger source of the slave
unit.
Push the -more- bezel button in the
Burst mode to display the second page.
8. Push the Source bezel button to select
External.
Confirm that Internal is selected in the
master and External is selected in the
slave for Source.
9. This screen shows signal outputs by
master-slave operation. The top wave-
form is the signal of the master unit.
10. When one meter long cables are used,
the slave signal(s) delays approxi-
mately 40 ns.
11. To make the delay between the master
and slave 0 (zero), set Delay of the
master unit.
12. The screen shows waveforms after
setting the delay to the master unit.
The top waveform is master and the
two bottom waveforms are slaves.
8
7
9
10
12
11