1502C MTDR User Manual B–1
Appendix B: Operator Performance ChecksThis appendix contains performance checks for many of the functions of the
1502C. They are recommended for incoming inspections to verify that the
instrument is functioning properly. Procedures to verify the actual performance
requirements are provided in the 1502C Service Manual.
Performing these checks will assure you that your instrument is in good working
condition. These checks should be performed upon receipt of a new instrument
or one that has been serviced or repaired. It does not test all portions of the
instrument to Calibration specifications.
The purpose of these checks is not to familiarize a new operator with the
instrument. If you are not experienced with the instrument, you should read the
Operating Instructions chapter of this manual before going on with these checks.
If the instrument fails any of these checks, it should be serviced. Many failure
modes affect only some of the instrument functions.
Item Tektronix Part Number
50 W precision terminator 011-0123-00
3-foot precision coaxial cable 012-1350-00
Disconnect any cables from the front-panel CABLE connector. Connect the
instrument to a suitable power source (a fully charged battery pack or AC line
source). If you are using AC power, make sure the fuse and power switch are
correct for the voltage you are using (115 VAC requires a different fuse than
230 VAC).
Pull the POWER switch on the front panel. If a message does not appear on the
display within a second or two, turn the instrument off. There are some failure
modes that could permanently damage or ruin the LCD if the power is left on for
more than a minute or so. Refer to Appendix C: Operator Troubleshooting in this
manual.
Option 05 instruments default to metric; however, you can change the metric
scale to ft/div in the Setup Menu or use the metric numbers provided. To change
the readings, press the MENU button. Using the n
o POSITION control, scroll
down to Setup Menu and press MENU again. Scroll down to Distance/Div is:
m/div and press MENU again. This will change to ft/div. Press the MENU button
repeatedly to return to normal operation mode. If the instrument power is turned