Performance Tests
TDS 500D, TDS 600B & TDS 700D Performance Verification and Specifications 1–87
HDisconnect the signal source from CH1, wait a few seconds, then
reconnect the signal.
HCONFIRM that the TRIG’D LED is lighted and the waveform is
stable.
HDisconnect all test equipment from the oscilloscope.
HPress save/recall SETUP, the main-menu button Recall Factory
Setup, and the side-menu OK Confirm Factory Init.