Performance Tests
TDS 500D, TDS 600B & TDS 700D Performance Verification and Specifications 1–63
d. Modify the initialized front-panel control settings:
HPress WAVEFORM OFF to turn off CH 1.
HPress CH 3 (AUX 1 on some TDS models) to display that channel.
HIf necessary, adjust the calibration generator output for 5 divisions of
amplitude. (With the 50W input of the TDS 794D, you may need to
double the output of the generator to get 5 divisions of amplitude.)
HSet the horizontal SCALE to 200 ms.
HPress SHIFT; then press ACQUIRE MENU.
HPress the main-menu button Mode; then press the side-menu button
Average.
HSelect 64 averages. Do this with the keypad or the general purpose
knob.
2. Confirm Main and Delayed Trigger outputs are within limits for logic levels:
a. Display the test signal:
HPress WAVEFORM OFF to turn off CH 3.
HPress CH 2 to display that channel.
HSet the vertical SCALE to 1 V.
HUse the vertical POSITION knob to center the display on screen.
b. Measure logic levels:
HPress MEASURE; then press the main-menu button Select
Measurement for Ch2.
HSelect high and low measurements. To do this, repeatedly press the
side-menu button –more– until High and Low appear in the side
menu (their icons are shown at the left). Press both side-menu
buttons High and Low.
c. Check Main Trigger output against limits:(Skip the first three subparts
of this substep for the TDS 794D only.)
HCHECK that the Ch2 High readout is 2.5 volts and that the Ch2
Low readout is 0.7 volts. See Figure 1–21.
HEnter high and low voltages on test record.
HPress VERTICAL MENU; then press the main-menu button
Coupling. Now press the side-menu button to toggle it to the
50 W setting.