Performance Tests
1–56 TDS 500D, TDS 600B & TDS 700D Performance Verification and Specifications
d. Check against limits: Do the following subparts in the order listed.
HCHECK that the Level readout in the side menu is within the limits
listed in Table 1–7.
Table 1–7: Delayed trigger level accuracy readout
Instrument Minimum Maximum
TDS 600B 9.863 V 10.137 V
TDS 500D/700D 9.8853 V 10.1147 V
TDS 794D 4.8953 V 5.1047 V
HEnter voltage on test record.
HPress the main-menu button Slope; then press the side-menu button
for negative slope. See icon at left. Press the main-menu button
Level. Repeat substep c.
HCHECK that the Level readout in the side menu is within the limits
listed in Table 1–7.
HEnter voltage on test record.
4. Disconnect the hookup:
a. First set the output of the DC calibration generator to 0 volts.
b. Disconnect the cable from the generator output at the input connector of
CH 1.