Performance Tests
1–16 TDS 500D, TDS 600B & TDS 700D Performance Verification and Specifications
Equipment RequiredThese procedures use external, traceable signal sources to directly check
warranted characteristics. The required equipment list follows this introduction.
Table 1–1: Test equipment
Item number and
description Minimum requirements Example Purpose
1. Attenuator,10X
(two required) Ratio: 10X; impedance 50 ; connec-
tors: female BNC input, male BNC
output
Tektronix part number
011-0059-02 Signal Attenuation
2. Attenuator, 5X Ratio: 5X; impedance 50 ; connec-
tors: female BNC input, male BNC
output
Tektronix part number
011-0060-02 Signal Attenuation
3. Adapter, BNC female-to-
Clip Leads BNC female-to-Clip Leads Tektronix part number
013-0076-00 Signal Coupling for Probe
Compensator Output Check
4. Terminator, 50 Impedance 50 ; connectors: female
BNC input, male BNC output Tektronix part number
011-0049-01 Signal Termination for
Channel Delay Test
5. Cable, Precision 50
Coaxial (two required) 50 , 36 in, male-to-male BNC
connectors Tektronix part number
012-0482-00 Signal Interconnection
6. Connector, Dual-Banana
(two required) Female BNC-to-dual banana Tektronix part number
103-0090-00 Various Accuracy Tests
7. Connector, BNC “T” Male BNC-to-dual female BNC Tektronix part number
103-0030-00 Checking Trigger Sensitivity
8. Coupler, Dual-Input Female BNC-to-dual male BNC Tektronix part number
067-0525-02 Checking Delay Between
Channels
9. Generator, DC Calibra-
tion Variable amplitude to ±104 V; accura-
cy to 0.1% Data Precision 82001Checking DC Offset, Gain,
and Measurement Accuracy
10. Generator, Calibration 500 mV square wave calibrator
amplitude; accuracy to 0.25% Tegam/Tektronix PG 506A
Calibration Generator1,2 To check accuracy of CH 3
Signal Out
11. Generator, Time Mark
(optional) Variable marker frequency from 10 ms
to 10 ns; accuracy within 2 ppm Tegam/Tektronix TG 501A
Time Mark Generator1,2 Checking Sample-Rate and
Delay-time Accuracy
12. Probe, 10X A P6139A, P6243, P6245, or P6339A
probe3Tektronix part number
P6139A or P6245 Signal Interconnection
13. 3.5 inch, 720 K or
1.44 Mbyte, DOS-com-
patible floppy disk
TDS Family Programmer Disk
(Tektronix part number
063–3002��00), included with
User Manual (Tektronix part
number 071–0130-XX)
Checking File System Basic
Functionality
14. Generator, Sine Wave 250 kHz to at least 500 MHz (higher
for higher-bandwidth scopes). Vari-
able amplitude from 60 mV to 2 Vp-p
into 50 . Frequency accuracy >2.0%
Rohde & Schwarz SMT or
SMY1,4 Checking Analog Bandwidth,
Trigger Sensitivity, Sample-
rate, External Clock, and
Delay-Time Accuracy