Saving Waveforms and Setups
TDS 684A, TDS 744A, & TDS 784A User Manual 3–131
Digitizing Oscilloscope (Rear Panel)
Controller
GPIB Connector
Figure 3–77: Connecting the Oscilloscope to a Controller
Select GPIB Port. To select the GPIB port, press SHIFT UTILITY
System (main) I/O (pop-up) Port (main) GPIB (pop-up).
Configure the GPIB Port. You must set two important GPIB parameters: mode
and address. To set those parameters:
Press SHIFT UTILITY System (main) I/O (pop-up) Port (main)
GPIB (pop-up) Configure (main) Talk/Listen Address, Hardcopy (Talk
Only), or Off Bus (side). (See Figure 3–78.)
Talk/Listen Address configures the port for controller-based system operation.
Use the general purpose knob or the keypad to define the address.
Hardcopy (Talk Only) configures the port for the hardcopy output without
controller supervision. Once so configured, the oscilloscope will send the
hardcopy data to any listeners on the bus when the HARDCOPY button is
pressed.
Pressing HARDCOPY with the port configured any other way causes an error,
and the oscilloscope responds with a message saying the selected hardcopy port
is currently unavailable.
Off Bus disconnects the oscilloscope from the bus.