Acquiring and Displaying Waveforms
3–4 TDS 684A, TDS 744A, & TDS 784A User Manual
Probe Compensated Correctly
Probe Overcompensated
Probe Undercompensated
Figure 3–1: How Probe Compensation Affects Signals
4. If you need to change the input impedance, press Coupling (main). Then
toggle the side menu selection W to get the correct impedance.
5. TDS 700A models only: Press SHIFT ACQUIRE MENU Mode (main)
Hi Res (side).
6. TDS 684A only: Press SHIFT ACQUIRE MENU Mode (main)
Average (side). Use the keypad to set Averages to 5.
7. Adjust the probe until you see a square wave with a perfectly flat top on the
display. Figure 3–2 shows where the adjustment is located.
Figure 3–2: P6139A Probe Adjustment