Adjustment Procedures
6–10 VITS100 NTSC VITS Inserter Instruction Manual
TSG/SPG
Test oscilloscope
Peak-to-peak
detector
OUTPUTINTPUT
BNC
Adaptor Peak-to-peak
detector head
Multiburst
Out
Figure 6–6: Peak-to-peak detector connections
8. Set the TSG/SPG Multiburst output as shown in Table 6–3.
Table 6–3: TSG/SPG Multiburst Settings
Control Setting
Frequency Range High
Markers On
Composite On
Amplitude Full
Sweep On
9. Connect a 75 W coaxial cable from the detector head output to the peak-to-
peak detector + input.
10. Enable peak-to-peak detector + input.
11. Set + input level to light the green LED (full clockwise).
12. Connect a 75 W coaxial cable from the peak-to-peak detector output to test
oscilloscope input.
13. Measure the sweep envelope. See Figure 6–7.