Reference
Manually Measuring Amplitude. The following procedure shows you how to manually measure the signal amplitude using both the graticule and the Voltage Cursors.
1.Perform the initial settings for Eye measurements described in Setup for Eye Measurements above.
2.Use the CLOCK BW soft key to select the 1 kHz filter. This setting reduces the effects of timing jitter, which can obscure the amplitude measurement.
3.To measure amplitude, use the graticule (1x gain) or use the voltage cursors. Ignore any overshoot or undershoot on the rising or falling edges.
NOTE. For Cursor measurements, you can use any gain setting, including variable gain, since the waveform and the cursors are equally affected by the gain setting. Higher gain settings help match the cursor to the waveform.
Signal sources should measure 800
Manually Measuring Aberrations. Serial sources should produce good signal transitions with a minimum of overshoot and ringing. To manually measure aberrations, perform the following steps:
1.Perform the steps in Setup for Eye Measurements on page
2.Position one voltage cursor at the peak of the overshoot at the top horizontal part of the waveform.
3.Position the second voltage cursor at the bottom of the top line of the waveform. Include any ringing (the oscillation following overshoot) in the measurement. Essentially, you are measuring the thickness of the top line of the waveform.
4.The Voltage Cursor readout provides the signal amplitude.
5.Perform the same
NOTE. Aberrations at the top or bottom line of the waveform should not exceed 10% of the signal amplitude since automatic equalizer circuits in receivers may be sensitive to larger aberrations.
3- 20 | WFM700 Series Waveform Monitors User Manual |