Figure 3- 15: XYZ pulse of Y channel with pass through selected
6.To measure horizontal timing, perform the following steps:
a.Switch to Input A and set the waveform monitor to one line sweep.
b.Use the horizontal position knob to set the SAV pulse to a major graticule tick mark, or use cursor mode and set a cursor on the SAV pulse.
c.Compare the timing to the signal on Input B by selecting Input B and adjusting the fine timing controls to match the timing position of Input A.