5 Maintenance

Model 3190

 

 

 

 

 

 

 

 

 

3.Unscrew the cell-retainer cap from the cell block by turning it counterclockwise until it is free.

Figure 5-2: Exploded View of MFC and Cell Block

4.Slowly withdraw the cap from the block. The cell should come out with the cap.

5.Carefully pull the cell off of the cap. DO NOT TOUCH THE SCREENED END OF THE CELL OR ANY FLUID THAT MAY BE LEAKING FROM IT.

6.Dispose of the cell in a safe manner, in accordance with all applicable ENVIRONMENTAL AND SAFETY laws.

5.2.4 Installing a Micro-Fuel Cell

To install a new Micro-Fuel Cell:

CAUTION: Do not scratch, puncture, or otherwise damage the sensing membrane of the Micro-Fuel Cell. If the membrane is damaged, the cell must be replaced.

5-4

Teledyne Analytical Instruments

Page 36
Image 36
Teledyne 3190 manual Installing a Micro-Fuel Cell, Exploded View of MFC and Cell Block