Teledyne 3190 manual Trace Oxygen Analyzer Installation

Models: 3190

1 42
Download 42 pages 34.02 Kb
Page 23
Image 23

Trace Oxygen Analyzer

Installation 3

 

 

 

 

 

 

 

 

 

fuse receptacle, to the right of the power terminal strip, holds a 0.5 A, very quick acting fuse. (See Fuse Replacement in chapter 5, Maintenance.)

The Power switch is located below the fuse receptacle.

WARNING: INSERT THE STRIPPED TIPS OF WIRES ENTIRELY INTO THE TERMINAL BLOCKS. DO NOT LEAVE EXPOSED WIRE OUTSIDE OF THE HOLES IN THE BLOCKS.

CAUTION: The control unit chassis must be isolated from the grounding system of the DC input power.

Analog Outputs: There are three DC output signal connectors with screw terminals on the panel. There are two wires per output with the polar- ity noted. See Figure 3-3. The outputs are:

0–10 V concentration:

Voltage rises with increasing oxygen concentra-

 

tion, from 0 V at 0 oxygen content to 10 V at full

 

scale oxygen content. (Full scale = 100 % of

 

programmed range.)

0–10 V Range ID:

03.33 V = Low Range, 06.66 V = High Range,

 

10 V = Air Cal Range.

4–20 mA concentration:

Current increases with increasing oxygen concen-

 

tration, from 4 mA at 0 oxygen content to 20 mA

 

at full scale oxygen content. (Full scale = 100 %

 

of programmed range.)

Alarm Relays: The three alarm-circuit connectors are screw terminals for making connections to internal alarm relay contacts. There is one set of contacts for each type of alarm. Contacts are Form C, with normally open and normally closed contact connections capable of switching up to 0.5 ampere at 125 V ac into a resistive load.

The alarm relay circuits are designed for failsafe operation, meaning the relays are energized during normal operation. If power fails the relays de- energize (alarms activated).

The contact connections are indicated diagrammatically on the rear panel as Normally Closed, Common, and Normally Open. Figure 3-2 explains how these act in failsafe operation.

Teledyne Analytical Instruments

3-5

Page 23
Image 23
Teledyne 3190 manual Trace Oxygen Analyzer Installation