Trace Oxygen Analyzer

Appendix

 

 

 

 

 

 

 

 

 

Overvoltage Category II

Appendix

A.1 Specifications

Ranges:

Two user selectable ranges can be set

 

between 0-10 ppm and 0-9,999 ppm

 

oxygen. Default ranges are 0-100 ppm and

 

0-1,000 ppm oxygen, and a 0-25 % (nomi-

 

nal) Air Calibration Range.

SignalOutput:

Voltage: 0–10 V dc, negative ground

 

Current: 4-20 mA, negative ground

Range ID:

0-10 V dc.

Display:

Lightemittingdiode(LED)display.

Alarms:

One high alarm relay, adjustable; one low

 

alarm relay, adjustable; one sensor failure

 

relay. (All are failsafe.)

System Operating Temp:

0-50 °C

Accuracy:

±2 % of full scale at constant temperature

 

±5 % of full scale through operating tem-

 

perature range (At 100 ppm and higher user

 

defined ranges) once temperature equilib-

 

rium is reached.

 

±1 ppm for 10 ppm range under above

 

conditions.

Response Time:

90 % in less than 65 seconds at 25 °C

 

(68 °F).

System Power Requirement:

AC (100 to 240 V ac, 47/440 Hz), or

 

DC (10-36 V dc); user specified.

System Enclosure:

Panel Mount: 2.81" H × 6.0" W × 2.87" D

 

(71.4 mm × 152.4 mm × 72.9 mm).

 

Face Plate: 3.75" H × 7.0" W

Teledyne Analytical Instruments

A-1

Page 39
Image 39