Trace Oxygen Analyzer

Maintenance 5

 

 

 

 

 

 

 

 

 

1.Disconnect the Power Source from the Control Unit.

2.Remove the new Micro-Fuel Cell from its protective bag.

3.Examine the O-ring at the base of the threaded portion of the cell- retainer cap, and replace it if it is worn of damaged.

3.Replace the cell on the end of cell-retainer cap, which is designed to fit snugly into the rim on the screen side of the cell.

4.Careful insert the cap and cell into the block, and screw the cap clockwise into the cell block until it is held firmly in the cell.

5.Reconnect the cell block electrical connector plug.

5.2.5 Cell Warranty Conditions

The Class A-2C, B-2C or Z-2C Micro-Fuel cell is used in the Model 3190. These cells are warranted for 6 months, with an expected life of 8 months from the date of shipment (under specified operating conditions—see Appendix). Note any Addenda attached to the front of this manual for special information applying to your instrument.

Note that the warranty period begins on the date of shipment. The customer should stock only one spare cell per instrument at a time. Do not attempt to stockpile spare cells.

If a cell was working satisfactorily, but ceases to function before the warranty period expires, the customer will receive credit toward the purchase of a new cell.

If you have a warranty claim, you must return the cell in question to the factory for evaluation. If it is determined that failure is due to faulty work- manship or material, the cell will be replaced at no cost to you.

Note: Evidence of damage due to tampering or mishandling will render the cell warranty null and void.