M703E Calibration and Verification M703E Calibrator Operator’s Manual

8.2.2. VIEWING O3 GENERATOR CALIBRATION POINTS

To view these calibration points, press:

SETUP X.X SECONDARY SETUP MENU
COMM VARS DIAG EXIT
Make sure that the M703E
is in standby mode.
SETUP X.X ENTER PASSWORD
818 ENTR
EXIT
Toggle these keys to enter
the correct PASSWORD
DIAG SIGNAL I/O
PREV NEXT ENTR EXIT
DIAG O3 GEN CALIBRATION
PREV NEXT ENTR EXIT
Continue pressing NEXT until ...
DIAG O3 GEN CALIBRATION
CAL PNTS EXIT
Toggle these keys to move
between calibration points
DIAG O3GEN 1) 500 MV, 5.0 MIN, ON
PREV NEXT INS DEL EDIT PRNT EXIT
STANDBY ACT =STANDBY
<TST TST> GEN STBY SEQ SETUP
SETUP X.X PRIMARY SETUP MENU
O3 SEQ CFG CLK PASS MORE EXIT