TELEDYNE API
M703E Calibrator Operator’s Manual Specifications, Approvals and Warranty

2. SPECIFICATIONS, APPROVALS AND

WARRANTY

2.1. SPECIFICATIONS

Table 2-1: M703E Analytical Specifications
Linearity +/- 1.0% of full scale
Precision 1.0 ppb
Stability +/- 2.0 ppb (photometer feedback mode)
Response Time 180 seconds to 95%
Stability (7-days) 1% photometer feedback; 3% without photometer feedback (CNST or REF)
Table 2-2: M703E Electrical and Physical Specifications
Temperature Range 5-40ºC
Humidity Range 0 - 95% RH, non-condensing
Dimensions (HxWxD) 7” (178 mm) x 17” (432 mm) x 24” (609 mm)
Operating Altitude 10,000 ft Maximum
Weight 35.5 lbs (16.1 kg) including internal zero air pump
AC Power 115VAC, 60Hz
230VAC,50HZ
Analog Outputs 1 user configurable output
Analog Output Ranges 0.1 V, 1 V, 5 V or 10 V
Range with 5% under/over-range
Analog Output Resolution 1 part in 4096 of selected full-scale voltage (12 bit)
Digital Control Outputs 12 opto-isolated outputs
Digital Control Inputs 12 opto-isolated outputs
Status Outputs 12 opto-isolated outputs, 5 defined, 7 spare
Serial I/O 2 ports: 1x RS-232; 1x RS-485 or RS-232 (configurable)
Communication speed: 300 - 115200 baud (user selectable)
Certifications
EN61326 (1997 w/A1: 98) Class A, FCC Part 15 Subpart B Section 15.107 Class
A, ICES-003 Class A (ANSI C63.4 1992) & AS/NZS 3548 (w/A1 & A2; 97)
Class A.
IEC 61010-1:90 + A1:92 + A2:95,
05744 Rev B 5