Portable Oxygen Analyzer Operation
Teledyne Analytical Instruments 17
2.1.2.1 V-MOUNT ADAPTER INSTALLATION
The V-Mount Adapter consists of a matching plastic plate with
integral V-grooves that attach to the rear of the instrument.
To install V-Mount Adapter remove battery compartment door by
prying up the hinged latch at the bottom of the cover, then slide the
adapter plate into grooves provided in rear case. Replace the battery
compartment door and secure door latch. See Figure 2-4.
Note: The door latch is a tight fit onto the battery cover. Use a
coin to gently pry up the latch.
Figure 2-4: V-Mount Adapter Installation
2.1.2.2 UNIVERSAL MOUNTING CLAMP INSTALLATION
The Universal Mounting Clamp is supplied with a 1/4-20 screw for
securing the clamp to the rear of the instrument. A threaded brass insert is
installed on the back of the instrument for this purpose. See Figure 2-5.
Figure 2-5: Brass Insert for Universal Mounting Clamp
2.1.3 Battery Installation
Note: Three “AA” alkaline batteries must be installed in the unit
before the analyzer will operate. The unit must be
recalibrated whenever new batteries are installed.
To install the batteries: