Portable Oxygen Analyzer Operation
Teledyne Analytical Instruments 29
Recalibrate after replacing the sensor.
Use properly installed alkaline batteries only.
Make sure the R17MED sensor is properly attached.
Remove the plastic flow diverter only when using the tee adapter.
Remove and save the plastic flow diverter when using the
sensor in non-flowing applications (incubators, tents, etc.)
Clean the case with isopropyl alcohol or mild detergent only.
– DON’T –
Use this analyzer if you suspect any malfunction.
Use the instrument in the presence of flammable gases.
Use anything but alkaline batteries.
Autoclave or freeze the sensor or instrument.
Open or try to repair a leaking or broken sensor.
Immerse the unit or sensor in any liquid.
Pass hot or cold gas mixtures over the sensor.
Adjust the reading in air after 100% calibration
Expose the unit to devices that produce high levels of radio,
short wave, microwave, x-ray, or high frequency
interference.
Use cleaning agents or liquids in the cable receptacles or
around the battery compartment.
Place the unit itself in a water vapor-saturated environment.
Expose the LCD to excessive sunlight.
Expose the unit to a condensing water environment such as a
mist tent.