Teledyne AX300-I manual Troubleshooting

Models: AX300-I

1 47
Download 47 pages 29.67 Kb
Page 36
Image 36

Service Manual

AX300-I

 

 

 

3.6.5.2CARE AFTER USE IN NITROUS OXIDE

CAUTION: THE R17MED SENSOR SHOULD NOT BE LEFT IN NITROUS OXIDE MIXTURES ANY LONGER THAN ABSOLUTELY NECESSARY.

After exposure to nitrous oxide mixtures, the sensor should be left in 100% oxygen overnight (e.g., left in a breathing circuit that has been flushed with pure oxygen). If this is not practical, when using the tee, remove the plastic flow diverter and leave the sensor in room air. If the oxygen reading continues to drop after each use in nitrous oxide the sensor should be removed from service. If the sensor can no longer be calibrated or if there is any sign of electrolyte leakage, the sensor should be disposed of in accordance with local regulations and the Material Safety Data Sheet (MSDS).

Note: The MSDS on this material is available upon request through the Teledyne Environmental, Health and Safety Coordinator. Contact at (626) 934-1592

3.7 Troubleshooting

The AX300-I oxygen analyzer provides a variety of built-in safety features that prevent its use when a fault is detected. When a unit displays the message √ SENSOR and sounds the audible and visual alarm continuously, it is an indication of a faulty connection between the sensor and the unit or an expired or faulty sensor. To determine where the difficulty lies, refer to the following guidelines in Table 3-2.

Table 3-2 Troubleshooting

Symptom

Why

What To Do

New sensor responds

If the sensor is new

A) Wait 1–2 hours and

slow or drifts.

and was just

recalibrate.

 

removed from its

 

 

sealed bag it may

 

 

need to run for

 

 

several hours.

 

Sensor will not read

Calibration in 100%

A) Recalibrate using dry gas

below 22 % after

was invalid or the

making sure the reading

calibration in 100% O2.

room air is

stabilizes before making

 

contaminated with

any adjustments.

 

excess oxygen.

B) Make sure that at least 6" (30

 

 

 

 

cm.) of tubing is attached to

 

 

the exhaust side of the tee

Teledyne Analytical Instruments

36

Page 36
Image 36
Teledyne AX300-I manual Troubleshooting