8.0 TROUBLESHOOTING GUIDE
Symptom | Possible Causes | Suggested Actions |
The LCD does not display. | No power to the instrument. | Check power supply to be |
|
| sure the correct voltage is |
|
| being supplied to the |
|
| transmitter. (13.5 to 50 |
|
| VDC) |
|
| Check polarity of transmitter |
|
| wiring. |
| Contrast level is set too low. | Set contrast level to a higher |
|
| value. Because the contrast |
|
| menu is always the top |
|
| menu, it can be reached by |
|
| pressing the upper MENU |
|
| SELECTION key at least 12 |
|
| times. Enter the calibrate |
|
| mode and increase the |
|
| value. |
| Sensor or signal conditioner | Check the sensor for proper |
| has a short that draws too | operation. To verify, |
| much power from the | disconnect all sensor wires |
| transmitter. | and cycle power off, then on. |
|
|
|
The pH fluctuates from | The slope calibration is mis- | Check buffers to make sure |
minimum scale to | calibrated. Typically the | they are the correct values. |
or the pH is locked at high or | transmitter is calibrated with |
|
low scale. | no slope (0 mV/pH). | Use the |
|
| calibration feature on the |
|
| Electrode Span Menu, then |
|
| perform a buffer calibration. |
|
| Check the electrode and |
|
| sensor for proper operation. |
| Wrong temperature | Examine the temperature |
| correction has been | and the Temperature |
| calibrated in the temperature | Calibration Menu. Adjust trim |
| trim. | as required. |
The pH or ORP reading short | The transmitter may have | Check buffers to make sure |
spans. | been | they are the correct values. |
|
| Use the |
|
| calibration feature on the |
Model