APPENDIX A - SPECIFICATIONS
MEASUREMENT RANGE
pH: 0.00 to 14.00, fully expandable and reversible, standard.
ORP: +1000 to
OUTPUT
POWER REQUIREMENTS
(with zero loop impedance)
Recommended | 24 vdc |
Maximum | 50 vdc |
Minimum | 13.5 vdc |
MAXIMUM LOOP IMPEDANCE (@ 24 vdc)
525 ohms for
OPERATING TEMPERATURE
DISPLAY
Menu driven, 32 character
ENCLOSURE
NEMA 4X, weatherproof
SHIPPING WEIGHT
Standard
ACCURACY
± 0.10% of full scale
LINEARITY
± 0.05% of full scale
SENSITIVITY
± 1.0 mV
STABILITY
± 0.2% per year @ 0°C to 70°C
RESPONSE TIME
1 second to reach 90% of the change.
REPEATABILITY
± 1.0 mV
TEMPERATURE COMPENSATION
pH: Automatic,
Specific pH compensation temperature correction) is available. Consult the factory.
ORP: Not used for compensation; however, temperature is displayed. Accuracy within ±0.1°C over a
50/60 Hz NOISE REJECTION Greater than 70 db
INPUT/OUTPUT ISOLATION
Maximum 300 volts between process input and any
CALIBRATION
Auto Buffer Calibration
Allows the definition of two buffer points, saved in memory, during the initial
Restart)
With 2 keystrokes, allows the technician to return the transmitter to a zero electrode offset (asymmetry potential) and to an ideal Nernstian slope (1.000 mV per ORP unit).
Model