T201 Ammonia Analyzer | Teledyne API Model T201 NH3 Analyzer Operator Manual |
1.1THEORY OF OPERATION
The
4NH3 + 5O2 4NO + 6H2O
| The resulting nitric oxide is then measured by the chemiluminescent | |
| reaction with ozone. Consult Section 10.1 in the T200 manual (provided | |
| with this instrument) for more details on the nitric oxide measurement. | |
| The analyzer uses two converters to oxidize the different sample gases. A | |
| high temperature catalytic converter, the M501 NH3, converts NH3 and | |
| NOx into NO creating the TNx channel. A second converter, housed inside | |
| the T201 analyzer and consisting of heated molybdenum, converts all of | |
| the NOx in the sample to NO producing the NOx channel. The nitric oxide | |
| channel is measured while bypassing both the M501 NH3 and | |
| molybdenum converter. A Nafion® drier operated in reflux mode, is | |
| installed prior to the molybdenum converter and the AutoZero valve. The | |
| drier removes ammonia and water from the switched stream. The | |
| ammonia would have otherwise been converted by the molybdenum and | |
| registered as NOx. This location provides the drier with continuous flow, | |
| thereby allowing it to stabilize faster. |
|
Note | Due to the Nafion® drier, the analyzer has a maximum ammonia range of | |
| 2.0 PPM. |
|
| The calculated gas concentrations, ammonia and nitrogen dioxide, are | |
| computed from the differences between the measured channels, as | |
| outlined in the following equations: |
|
| NO (Nitric Oxide) Concentration = Measured |
|
| TNx (Total Nitrogen) Concentration = Measured (NH3 + NO2 + NO) | |
| NOx (Total Nitrogen – Ammonia) Concentration = | Measured |
| (NO2 + NO) |
|
| NH3 (Ammonia) Concentration = Calculated as follows: |
|
| [(TNx - NOx) / (NH3_CE_FACTOR1)] |
|
| NO2 (Nitrogen Dioxide) Concentration = Calculated as follows: | |
| [(NOx – NO) / (NO2_CE_FACTOR1)] |
|
Note | The nitrogen dioxide efficiency factor (NO2_CE_FACTOR1) must be | |
| calculated by conducting a GPT (gas phase titration) as described in Section |
8.3.3 of the T200 manual and Section 6.0 of this manual. This value gives a good indication of how well the molybdenum is converting nitrogen dioxide into nitric oxide. In a properly functioning analyzer the value should be close to 1.0. It is recommended that the molybdenum efficiency be checked every three months.
10 | 07271B DCN6646 |