Calibration Procedure | Teledyne API Model T201 NH3 Analyzer Operator Manual |
2.1ZEROING THE ANALYZER
The analyzer can be zeroed by either applying zero air straight to the sample port or through the optional Zero/Span valves, if installed. It is important to remember that if the analyzer was previously sampling ammonia gas, prior to conducting a Zero calibration, it will take some period of time before the zero baseline is reached. The process of zeroing the analyzer consists of delivering dry zero air to the analyzer. The operator then manually zeros the TNx and NOx channels individually through the user interface. The following Tables outline the steps necessary to zero the analyzer.
Table
Important: Each channel (TNx and NOx) must be zeroed individually
Step No. | Action | Comment |
|
|
|
1. | Press CAL | The T201 enters the calibrate mode from sample |
|
| mode. Confirm zero gas is flowing past the sample |
|
| port. |
2. | Channel Selection | Press TNX or NOx then Press ENTR. |
3. | Range Selection | Press LOW or HIGH range then Press ENTR. |
|
| Always calibrate the LOW range, section 3.0 below. |
4. | NH3 STB Reading | Wait for the displayed stability reading to fall below |
|
| 1.0 PPB. The lower the stability reading the better the |
|
| zero value. |
5. | Press ZERO | If you change your mind after pressing ZERO, you |
|
| can still press EXIT without zeroing the instrument. |
|
| You don’t have to enter concentration values. The |
|
| analyzer knows to apply 0 PPB. |
6. | Press ENTR | Pressing ENTR actually adjusts the Offset value for |
|
| the selected channel. |
7. | Press EXIT | The T201 returns to sampling mode. |
8. | Check | All displayed concentrations should read zero. The |
| Concentrations | offset values should be close to zero |
| /Offsets |
|
16 | 07271B DCN6646 |