Teledyne T201 manual Analog Output Pin-Outs

Models: T201

1 44
Download 44 pages 48.34 Kb
Page 28
Image 28

Configurable Analog Output Overview Teledyne API Model T201 NH3 Analyzer Operator Manual

Table 3-2. Analog Output Pin-Outs

 

PIN

 

ANALOG OUTPUT

VOLTAGE SIGNAL

CURRENT SIGNAL

 

 

1

 

A1

V Out

I Out +

 

 

2

 

Ground

I Out -

 

 

 

 

 

 

 

 

 

 

 

 

 

3

 

A2

V Out

I Out +

 

 

 

 

 

 

 

 

4

 

Ground

I Out -

 

 

 

 

 

 

5

 

A3

V Out

I Out +

 

 

6

 

Ground

I Out -

 

 

 

 

 

 

7

 

A4

V Out

N/A

 

 

 

 

 

 

 

 

8

 

Ground

N/A

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Additionally A1, A2 and A3 may be equipped with optional 0-20 mA, or

 

 

4-20, current loop drivers. The 4-20 mA option is not available on A4.

Note

In actuality the analog output configuration of the analyzer may be different

 

 

than stated above. The outputs can be configured differently at the factory

 

 

depending on whether they were assigned during the procurement of the

 

 

product. It is possible to check the configuration of the analyzer by accessing

 

 

the ANALOG I/O CONFIGURATION through the DIAG menu.

The analyzer operates in Dual Mode during gas detection. This means each of the measured concentrations can have two separate slopes and offsets, one for the low range and one for the high range. Though uncommon, a user may decide to calibrate the analyzer with nitric oxide at, for example, 100 PPB using the LOW range and then perform another calibration at 450 PPB using the HIGH range. Through the analog outputs the user can then assign analog output A1 to TNxCNC1 and output A2 to TNxCNC2. Gases with the “1” designation use the slope and offset for the LOW range, while gases with the “2” designation will use the slope and offset for the HIGH range. It is recommended that both the LOW and HIGH ranges be calibrated at the same time. Independent of whether the HIGH range is actually being outputted to the analog outputs.

For analog output configuration, signal type and range selection and calibration, use the T200 manual.