Humidity Sensor Calibration

Press select a fifth time; the unit displays 0, 1, 2, or 3 in the rainfall window. 2 is the normal display. Press the maximum or minimum key to enter the proper number (see calibration reference for more information).

Last

Hold

Daily

2

IInch

Monthly

Term

Press select a sixth time; the unit will display a reading ranging from .01 to .1 in the rainfall window. Press the maximum or minimum button to enter the proper number which is usually .01.

Last

Hold

Daily

.01

IInch

Monthly

Term

Press select a seventh time - the unit displays wind direction. Press the maximum or minimum key to enter the actual wind direction.

Press select an eighth time - the unit will return to the normal display mode.

Humidity Sensor Calibration

This function is only used if a new humidity sensor is installed. Press select and humidity hold to enter this mode. The first number to appear is the voltage at 0% humidity that the sensor outputs. Press select again and a second number will appear. This larger number is the voltage that the sensor outputs at 75% humidity. The humidity calibration numbers are found on the humidity/temperature sensor and on the back of the instrument console. These calibration numbers are supplied to us by the manufacturer of the sensor and are normally not changed. If the unit is in protect mode this function is disabled.