5.Cut all of the cable ties coming from the tube holder following the two wires from the tube holder to where it terminates. Note: there will be two different termination points for each tube holder. Once this has been located, cut the wires (marking them to ensure easy replacement).
6.The tube holder(s) can now be removed by pushing the holder into the main chamber and feeding the wires through the
7.Replace tube holder(s). (Part number
8.In order to refit the tube holder, reverse the above procedure using butt crimps to reconnect the wires.
Note: If more than one tube holder is to be replaced, the wiring schematic diagram may be of assistance.
Replacement of starters
1.Locate the starter to be changed. These are positioned in a vertical row on the right hand side of the internal chassis assembly.
2.Turn starter(s) 30 degrees
3.Reverse above procedure with recommended starter.
Replacement of starter holder
1.Locate the starter holder to be changed. It is located on the right hand side of the internal chassis assembly.
2.Lever off the starter holder using a flat bladed screwdriver. The holder should now be loose.
3.Replace unit (part number
4.To replace the starter holder first push the center of the two black spigots, which hold the starter holder (x2) away from you i.e. towards the crosslinker chamber. As the reflector has been loosened, this should be done fairly easily with a small,
5.The starter can now be relocated.