UV CROSSLINKERS

UVP

The new setting will appear on the LED display.

Push ENTER on the touchpad.

The new setting is now installed.

B.Preset ultraviolet time exposure

Push PRESET and then TIME on the tactile touchpad. The red light at each position should now be lit and the preprogrammed UV exposure time of 2 minutes should be displayed in the LED.

Push START on the touchpad. After a slight delay to energize the UV tubes, the LED will begin to countdown. NOTE: THE TIME EXPOSURE IS SET IN MINUTES AND TENTHS OF A MINUTE.

The unit will automatically stop at the end of the exposure cycle and will beep five times. Exposure is now complete.

Information: Though the PRESET ULTRAVIOLET TIME EXPOSURE is factory set to 2 minutes, it is possible to change this if necessary, as follows:

Push and hold the PRESET on the touchpad until you hear an audible signal, then push TIME.

Set your new PRESET UV TIME EXPOSURE by pushing the numbers on the tactile touchpad. The new setting will appear on the LED display.

Push ENTER and the new setting will be installed.

C. User set UV energy exposure

Sometimes it may be necessary for you to set your own standards for exposure. This is easily accomplished as follows:

Push the ENERGY on the tactile touchpad then set your energy exposure requirements by pushing the numbers on the touchpad.

The energy exposure settings should now be displayed on the LED in flashing mode.

- 4 -

REMEMBER: YOUR ENERGY EXPOSURE SETTINGS DISPLAYED MUST BE MULTIPLIED BY 100. If settings are correct, push ENTER on the touchpad.

Page 6
Image 6
UVPS 81-0112-01 Rev C manual UV Crosslinkers UVP