Quantum Data 700 Series manual Calibrating Outputs

Page 10

Calibrating

Calibrating Outputs

1.Select Options > Calibrate > Reset to set the generator’s calibration factors to default nominal values.

2.Press Options > Calibrate > Full Scale.

3.Set the voltmeter’s scale factor to be able to measure a nominal 1000 millivolts DC.

4.Using a VGA-to-BNC cable, connect the generator’s red output to the positive input on the voltmeter, with a 75 ohm (+/- 1%) input terminator. Connect the negative lead of the meter to any ground pin.

5.Select R_FS+ or -R_FSto set the DC output level to 1000 mV +/- 3 mV.

6.Repeat steps 4 and 5 for the green and blue outputs.

7.Select Save.

8.Select Options > Calibrate > Zero.

9.Connect the red output to the positive input of the voltmeter.

_____ Calibrate

_

 

_Full_Scale____

Zero

 

_Reset

 

 

 

 

 

 

 

 

 

 

 

 

 

 

_R_FS+_____

___-R_FS

 

_G_FS+_______

_-G_FS

 

_B_FS+_______

_-B_FS

 

_Save_________

_Back

 

 

 

 

 

 

 

 

 

 

_R_Zero+____

-R_Zero

 

_G_Zero+____

-G_Zero

 

_B_Zero+____

-B_Zero

 

_Save__________

Back

 

 

 

10.Select R_Zero+ or -R_Zeroto set the DC output level to 0 mV +/- 3 mV.

11.Repeat steps 9 and 10 for the green and blue components.

12.Select Save.

13.Select Options > NTSC_CAL.

14.Adjust the oscilloscope's timebase and sweep trigger to view the entire waveform. Adjust the G_NTSC so that its chrominance is 100 IRE or 714mV.

15.Using a vector scope, use the R_NTSC and B_NTSC options to adjust the red and green outputs using so that their signals align with the 2% boxes.

16.Press Save.

Note: The modulated RF output does not have any calibration settings.

10

Image 10
Contents Page Contents Using Battery Power Using AC PowerGetting Started Computer Interfaces DVI-ILCD Display User InterfaceFunction Buttons Selecting Formats Viewing Formats by Signal TypeSelecting Formats Reloading Standard FormatsSelecting Images Selecting ImagesGating Outputs Setting Power Save Preference Setting Power Save PreferencesOptions Menu Calibrating Equipment Needed for CalibrationCalibrating Outputs Locking Calibration Settings Adjusting Horizontal Scanning FrequencyDetecting Formats Supported by a Monitor Learn MonitorTest Menu Using Test SequencesRunning Test Sequence Manually Running Test Sequence in Burn-In Mode Creating Test Sequence for DDC DisplaysTesting Hdcp Interfaces Using Generator with CS-1 ProbeTesting Audio Special TestsGenerating Pseudo Random Noise About VGM Using Video Generator ManagerVGM Quick Tour Creating Video Formats Editing Video FormatsModifying Test Sequence Editing the Test SequenceGrid Pattern DescriptionsFlatWht/R,G,B RevGridWindow RasterCutoff FocusDot12 RegulateChecker SMPTEBarPulseBar RampBurst Updating Firmware Updating FirmwareSpecifications Video FormatsService and Support Contacts