45
Users Manual
DC Voltage
Range
300 mV
3V
30V
300V
1000 V
100mV
1000 mV
10 V
100V
1000 V
Slow
—
—
—
—
—
1µ V
10µ V
100µ V
1mV
10mV
Resolution
Medium
10µ V
100µ V
1mV
10mV
100mV
—
—
—
—
—
Fast
100µ V
1mV
10mV
100mV
1V
—
—
—
—
—
Accuracy
(6 Months) | (1 Year) |
|
|
002 % + 2 | 0.025 % + 2 |
0.02 % + 2 | 0.025 % + 2 |
0.02 % + 2 | 0.025 % + 2 |
0.02 % + 2 | 0.025 % + 2 |
0.02 % + 2 | 0.025 % + 2 |
|
|
0.02 % + 6 | 0.025 % + 6 |
0.02 % + 6 | 0.025 % + 6 |
0.02 % + 6 | 0.025 % + 6 |
0.02 % + 6 | 0.025 % + 6 |
0.02 % + 6 | 0.025 % + 6 |
|
|
Input Impedance
10 Me in parallel with <100 pF
Note
In the dual display mode, when the volts ac and volts dc functions are selected, the 10 MΩ dc input divider is in parallel with the 1 MΩ ac divider.
Normal Mode Rejection Ratio
>80 dB at 50 Hz or 60 Hz, slow and medium rates
>54 dB for frequencies between
>60 dB at 50 Hz, fast rate (Note: Fast rate has no filtering)
Maximum Allowable AC Voltage While Measuring DC Voltage or (AC + DC) Voltages
Range | Max Allowable Peak AC |
|
| Voltage |
|
|
|
Peak Normal Mode Signal
300 mV | 100 mV | 15 V |
3 V | 1000 mV | 15 V |
30 V | 10 V | 1000 V |
300 V | 100 V | 1000 V |
1000 V | 1000 V | 1000 V |
|
|
|
* NMRR is the Normal Mode Rejection Ratio
† Normal Mode Rejection Ratio at 50 Hz or 60 Hz ± 0.1 %
NMRR* >80 dB†
15V
15V
50V
50V
200V
NMRR >60 dB†
15V
15V
300V
300V
1000 V
Common Mode Rejection Ratio
>90 dB at do, 50 or 60 Hz, (1 ke unbalanced, medium and slow rates)