Self-Test Diagnostics and Error Codes

45

Users Manual

3.Carefully remove the fuse and replace it with one rated as listed above.

4.Reverse the disassembly procedure to reassemble the meter.

Self-Test Diagnostics and Error Codes

If the meter fails the self-test diagnostics performed at power-up, an error code is displayed in the primary display and "ERROR" is displayed in the secondary display. An error code is displayed for two seconds or until a button is pressed. Error codes are listed in Table 6-1.

Error No.

1

2

3

4

5

6

7

8

9

Table 6-1. Self-Test Error Codes

Meaning

ROM test failed

External RAM test failed

Internal RAM test failed

Display self-test failed

Display dead

EEPROM instrument configuration corrupted

EEPROM calibration data corrupted ("UNCAL" annunciator also lights)

A/D chip dead

Measurement self-test failed

Performance Tests

The meter should be calibrated and in operating condition when you receive it.

The following performance tests are provided to ensure that the meter is in proper operating condition. If the instrument fails any of the performance tests, then calibration adjustment and/ or repair is needed. To perform these tests, you will need a Fluke 5700A Multifunction Calibrator and 5725A Amplifier (or equivalents).

Each of the measurements listed in the following steps assume the instrument is being tested after a one-hour warmup, in an environment with an ambient temperature of 18 °C to 28 °C, and a relative humidity of less than 90 % (70 % for 1,000 krange and above).

Note

All measurements listed in the performance test tables are made in the medium reading rate unless otherwise noted.

1.Power-up the meter and allow it to stabilize for one hour.

2.Connect a cable from the Output VA HI and LO connectors of the 5700A to the Xand COM connectors on the Fluke 45.

Select the function and range on the Fluke 45 and the input level from the 5700A using the values listed in Table 6-2. The display should read between the minimum and maximum values listed in the table.

3.Connect a cable from the Output VA HI and LO connectors of the 5700A to the 100 mA and COM connectors on the Fluke 45. Select the function and range on the Fluke 45 and the input level from the 5700A using the values contained in Table 6-3. The display should read between the minimum and maximum readings listed in the table.