45
Users Manual
3.Carefully remove the fuse and replace it with one rated as listed above.
4.Reverse the disassembly procedure to reassemble the meter.
Self-Test Diagnostics and Error Codes
If the meter fails the
Error No.
1
2
3
4
5
6
7
8
9
Table
Meaning
ROM test failed
External RAM test failed
Internal RAM test failed
Display
Display dead
EEPROM instrument configuration corrupted
EEPROM calibration data corrupted ("UNCAL" annunciator also lights)
A/D chip dead
Measurement
Performance Tests
The meter should be calibrated and in operating condition when you receive it.
The following performance tests are provided to ensure that the meter is in proper operating condition. If the instrument fails any of the performance tests, then calibration adjustment and/ or repair is needed. To perform these tests, you will need a Fluke 5700A Multifunction Calibrator and 5725A Amplifier (or equivalents).
Each of the measurements listed in the following steps assume the instrument is being tested after a
Note
All measurements listed in the performance test tables are made in the medium reading rate unless otherwise noted.
1.
2.Connect a cable from the Output VA HI and LO connectors of the 5700A to the Xand COM connectors on the Fluke 45.
Select the function and range on the Fluke 45 and the input level from the 5700A using the values listed in Table
3.Connect a cable from the Output VA HI and LO connectors of the 5700A to the 100 mA and COM connectors on the Fluke 45. Select the function and range on the Fluke 45 and the input level from the 5700A using the values contained in Table