Fluke 45 user manual DC Current, AC Current, Appendices, Maximum Input

Models: 45

1 118
Download 118 pages 5.63 Kb
Page 106
Image 106
DC Current

Appendices

Specifications A

DC Current

Range

Slow

30 mA

 

100 mA

 

10 A

 

10 mA

100 nA

100 mA

1 µ A

10 A

100

µ A

 

 

 

* Typical at full range

Resolution

Medium

1µ A

10µ A

1mA

Fast

10µ A

100µ A

10mA

Accuracy

0.05% + 3

0.05% + 2

0.2% + 5

0.05% +

50.05% + 5

0.2% + 7

Burden Voltage

0.45V

1.4V

0.25V

0.14V

1.4V

0.25V

Maximum Input

To be used in protected, low energy circuits only, not to exceed 250 V or 4800 Volt-Amps. (IEC 664 Installation Category II.)

mA

300 mA dc or ac rms. Protected with a 500 mA, 250V, IEC 127-sheet 1, fast

 

blow fuse

A10 A dc or ac rms continuous, or 20 A dc or ac rms for 30 seconds maximum. Protected with a 15 A, 250 V, 10,000 A interrupt rating, fast blow fuse.

Note

Resistance between the COM binding post and the meter’s internal measuring circuits is approximately .003 .