Appendices
Specifications A
DC Current
Range
Slow
30 mA | — |
|
100 mA | — |
|
10 A | — |
|
10 mA | 100 nA | |
100 mA | 1 µ A | |
10 A | 100 | µ A |
|
|
|
* Typical at full range
Resolution
Medium
1µ A
10µ A
1mA
—
—
—
Fast
10µ A
100µ A
10mA
—
—
—
Accuracy
0.05% + 3
0.05% + 2
0.2% + 5
0.05% +
50.05% + 5
0.2% + 7
Burden Voltage
0.45V
1.4V
0.25V
0.14V
1.4V
0.25V
Maximum Input
To be used in protected, low energy circuits only, not to exceed 250 V or 4800
mA | 300 mA dc or ac rms. Protected with a 500 mA, 250V, IEC |
| blow fuse |
A10 A dc or ac rms continuous, or 20 A dc or ac rms for 30 seconds maximum. Protected with a 15 A, 250 V, 10,000 A interrupt rating, fast blow fuse.
Note
Resistance between the COM binding post and the meter’s internal measuring circuits is approximately .003 Ω .
AC Current
Range
Slow
10 mA |
| 100 nA |
30 mA |
| — |
100 mA |
| 1 µ A |
10 A |
| 100 µ A |
|
|
|
* Typical at full range |
| |
|
|
|
Resolution
Medium
—
1µ A
10µ A
1mA
Fast
—
10µ A
100µ A
10mA
Burden
Voltage*
0.14V
0.45V
1.4V
0.25V