Fluke 45 True RMS AC Voltage, AC-Coupled, Appendices, Maximum Input, Accuracy, Decibel Resolution

Models: 45

1 118
Download 118 pages 5.63 Kb
Page 104
Image 104
Appendices

Appendices

Specifications A

Maximum Input

1000V dc or peak ac on any range

True RMS AC Voltage, AC-Coupled

Range

300mV

3V

30V

300V

750V

100mV

1000 mV

10V

100V

750V

Slow

1µ V

10µ V

100µ V

1mV

10mV

Resolution

Medium

10µ V 100µ V

1mV

10mV

100mV

Fast

100µ V

1mV

10mV

100mV

1V

Accuracy

 

Linear Accuracy

 

Frequency

 

 

 

Slow

Medium

Fast

 

 

 

 

 

20-50 Hz

1 % + 100

1 % + 10

7 % + 2

50 Hz-10 kHz

0.2 % + 100

0.2 % + 10

0.5 % + 2

10-20 kHz

0.5 % + 100

0.5 % + 10

0.5 % + 2

20-50 kHz

2 % + 200

2 % + 20

2 % + 3

50-100 kHz

5 % + 500

5 % + 50

5 % + 6

 

 

 

 

dB Accuracy

Slow/Med

 

Fast

 

 

 

 

 

 

 

0.150.72

0.080.17

0.110.17

0.290.34

0.700.78

Power*

2 % + 10

0.4% + 10

1 % + 10

4 % + 20

10 % + 50

Max

Input at

Upper

Freq

750V

750V

750V

400V

200V

* Error in power mode will not exceed twice the linear accuracy specification

Accuracy specifications apply within the following limits, based on reading rate:

Slow Reading Rate: Between 15,000 and 99,999 counts (full range)

Medium Reading Rate: Between 1,500 and 30,000 counts (full range)

Fast Reading Rate: Between 150 and 3,000 counts (full range)

Decibel Resolution

Resolution

Slow & Medium

Fast

0.01 dB

0.1 dB