Fluke 45 The *RDT and *RAT? commands are not implemented, There are no additional data structures

Models: 45

1 118
Download 118 pages 5.63 Kb
Page 118
Image 118
The *RDT and *RAT? commands are not implemented.

45

Users Manual

18.The size of the resource description, if the *RDT command or *RDT? query are implemented, Sections 10.30 and 10.31.

The *RDT and *RAT? commands are not implemented.

19.A description of the states affected by *RST (Section 10.32), *LRN? (Section 10.17), *RCL (Section 10.29), and *SAY (Section 10.33).

*LRN.2, *RCL, and *SAT are not implemented. *RST restores the meter to the state assumed at power-up (using the stored configuration), except for those items specifically forbidden by the *RST command definition.

20.A description of the scope of the self-test performed by the *TST? query, Section 10.38.

*TST? performs the tests listed under "*TAT?" in Table 5-8 of the Users Manual. The meter reverts to the stored configuration after performing these tests.

21.A description of additional status data structures used in the device’s status reporting, Section 11.

There are no additional data structures.

22.For each command, a statement describing whether is overlapped or sequential. All commands are sequential; none are overlapped.

23.For each command, the device documentation shall specify the functional criteria that are met when an operation complete message is generated in response to that command, Section 12.8.3.

Operation complete is generated when the command is parsed. Note that MEAS?, MEAS1?, and MEAS2? prevent the parser from continuing until all requested measurements have completed. (This includes any time waiting for the trigger). The VAL, VAL2?, and VAL2? queries will also wait for trigger if the needed display measurements have never been done

Page 118
Image 118
Fluke 45 user manual The *RDT and *RAT? commands are not implemented, There are no additional data structures