Fluke 45 Frequency, Sensitivity of AC Voltage, Sensitivity Level of AC Current, Input Protection

Models: 45

1 118
Download 118 pages 5.63 Kb
Page 109
Image 109
Frequency

45

Users Manual

Open Circuit Voltage

3.2V maximum Continuity Capture Time

50 us maximum, 10 us typical

Input Protection

500 volts dc or rms ac

Note

When the meter is set to measure frequency and there is no input signal (i.e., input terminals are open), the meter may read approximately 25 kHz (rather than the expected zero). This is due to internal capacitive pickup of the inverter power supply into the high-impedance, input circuitry. With source impedance of <2 k, this pickup will not affect the accuracy or stability of the frequency a reading.

Frequency

Frequency Range

5 Hz to >1 MHz

Applicable Functions

Volts ac and Current AC

Range

Resolution

Accuracy

1000 Hz

10kHz

100kHz

1000 kHz

1MHz*

Slow & Medium

.01 Hz

.1 Hz

1Hz

10Hz

100Hz

Fast

.1 Hz

1Hz

10Hz

100Hz

1kHz

05% + 2

.05% + 1

.05% + 1

.05% + 1 Not Specified

* Specified to 1 MHz, but will measure above 1 MHz.

Sensitivity of AC Voltage

Frequency

5Hz-100 kHz

100kHz - 300 kHz

300kHz - 1 MHz Above 1 MHz

Level (sine wave)

30 mV rms

100 mV rms

1 V V rms

Not specified

Sensitivity Level of AC Current

Frequency

Input

 

 

5 Hz-20 kHz

100 mA

45 Hz-2 kHz

10 A

 

 

Level

>3 mA rms

>3 A rms