Fluke 45 user manual 4.Input Terminals

Models: 45

1 118
Download 118 pages 5.63 Kb
Page 28
Image 28
Input Terminals

45

Users Manual

Neither function modifiers (REL, dB, HOLD, and MN MX) nor the manual range mode can be selected in the secondary display. Measurement ranges in the secondary display are always selected through autoranging.

Input Terminals

The input terminals, shown in Figure 3-4, are located on the left of the front panel.

The meter is protected against overloads up to the limits shown in Table 3-1. Exceeding these limits poses a hazard to both the meter and operator.

Volts, Ohms, Diode Test Input Terminal

V

10A

 

600V CAT I

!

1000V CAT I

 

COM Figure 3-4. Input Terminals 100 mA

FUSE F1

500 mA F 250V

FUSED

Amperes Input Terminal. For Current Measurements up to 10A continuous (or 20A for 30 sec).

Milliamperes Input Terminal. For Current Measurements up to 100mA.

 

 

 

 

45 DUAL DISPLAY MULTIMETER

 

 

 

V

 

10A

 

 

 

 

 

 

 

 

 

 

 

 

 

 

600V CAT I

!

 

 

 

 

 

 

CAL

1000V CAT I

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

ENABLE

COM

 

100

 

 

 

 

 

 

 

mA

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

FUSE F1

V

A

 

 

REL

dB

2ND

 

500 mA

 

 

 

 

 

 

 

 

 

 

 

F 250V

 

 

 

 

REF#

REF

LOCAL

FUSED

V

A

FREQ AUTO

 

 

 

 

 

 

 

HOLD

MN MX

RATE

 

 

 

 

 

 

 

 

 

 

COMP

HI

LO THRESH

ADDR

BAUD

POWER

Note: Frequency Measurements are Normally taken

from the VManual background Input, although the 10A and 100mA Inputs can be used.

Common Terminal. Return Terminal for all Measurements.

aam12f.eps

Figure 3-4. Input Terminals

3-4