45
Users Manual
Table
Range
300mV
3V
30V
300V
750V
Maximim Input Voltage
1 V rms
6 V rms
60 V rms
750 V rms
750 V rms
The input signal sensitivity is listed under the frequency specifications in Appendix A. These values are based on sine waveforms. The signal level must be increased for lower crest factor inputs (the crest factor is the ratio of the peak voltage to the ac rms voltage of the waveform). If the input signal is below the required level, the frequency will be displayed as zero. If the measurements are unstable, the input signal may be near the threshold level.
Selecting A Function Modifier
(K, I, H, J)
Selecting a function modifier (see Figure
To use a function modifier, press a function button to select a primary function, then press a function modifier button (or buttons). Modified readings are shown only on the primary display.
After a function modifier has been selected, pressing any (white) function button turns off all modifiers, causes the secondary display to go blank, and returns unmodified readings to the primary display.