General
Common Mode Voltage | 1000 V dc or peak ac maximum from any input to earth |
Size | 9.3 cm high, 21.6 cm wide, 28.6 cm deep (3.67 in high,8.5 in wide, |
| 11.27 in deep) |
Weight | Net, 2.4 kg (5.2 Ibs) without battery; |
| 3.2 kg (7.0 Ibs) with battery; |
| Shipping, 4.0 kg (8.7 Ibs) without battery; |
| 4.8 (10.5 Ibs) with battery |
Power | 90 V to 264 V ac (no switching required), 50 Hz and 60 Hz < 15 VA |
Standards | maximum |
| |
| Complies with: IEC 348, UL1244, CSA Bulletin 566B |
EMC: Part 15 subpart J of FCC Rules, and VDE 0871. | |
| |
| Baud rates: 300, 600,1200,2400,4800 and 9600 |
| Odd, even or no parity |
| One stop bit |
Options
Battery (Option | Type | 8 V, |
| Operating Time | 8 hours (typical). N lights when less than |
|
| 1/2 hour of battery operation remains. |
|
| Meter still meets specifications. |
| Recharge Time | 16 hours (typical) with meter turned off |
|
| and plugged into line power. Battery will |
|
| not charge when meter is turned on. |
Capability codes | SH1, AH1, T5, L4, SRI, RL1, PP0, DC1, | |
|
| DT1, E1, TED, LEO and C0 |
| External Trigger Input |
|
| VIH | 1.35 V minimum |
| VIL | 1.25 V maximum |
| Input Threshold Hysteresis | 0.6 V minimum |