45
Users Manual
Accuracy
RangeFrequency
Slow
Accuracy
Medium
Fast
mA (To 100 mA) mA (To 100 mA) mA (To 100 mA)
A
A (0.5 to 1A) A (0.5 to 1A)
50
10
50
2 % + 100
0.5% + 100
2 % + 200
2 % + 100
1 % + 100
2 % + 300
1 % + 300
2 % + 10
0.5% + 10
2 % + 20
2 % + 10
1 % + 10
2 % + 30
1 % + 30
7 % + 2
0.8% + 2
2 % + 3
7 % + 2
1.3% + 2
7 % + 4
1.3% + 4
mA accuracy specifications apply within the following limits, based on reading rate:
Slow Reading Rate: | Between 15,000 and 99,999 counts (full range) |
Medium Reading Rate: | Between 1,500 and 30,000 counts (full range) |
Fast Reading Rate: | Between 150 and 3,000 counts (full range) |
|
|
Maximum Crest Factor
3.0
Maximum Input
To be used in protected, low energy circuits only, not to exceed 250 V or 4800
mA | 300 mA dc or ac rms. Protected with a 500 mA, 250 V, IEC |
| blow fuse |
A10 A dc or ac rms continuous, or 20 A dc or ac rms for 30 seconds maximum. Protected with a 15 A, 250 V, 10,000 A interrupt rating, fast blow fuse.
Note
Resistance between the COM binding post and the meter’s internal measuring circuits is approximately .003Ω .