Maintenance 6
Current Input Fuses
If the fuse is good, the meter will read between .04 Ω and 1.0 Ω . If the fuse is blown, the meter will read >10 MΩ to OL.
Replacing the 100 mA Input Fuse
WWarning
For protection against fire or arc flash, replace a blown fuse only with one of identical rating.
The 100 mA input fuse is mounted in the front panel input jack (see Figure
To replace this fuse, first unplug the line cord. Then press in on the input jack and turn it 90 degrees
Replace a blown fuse with one of identical rating (see Table
F1 Fuse (500 mA, 250V, Fast Blow)
1500 A Minimum Breaking Capacity
| Front Panel Input Terminal | |||||
V |
|
|
|
|
|
|
600V |
|
|
|
|
|
|
1000V CAT |
|
|
| 10A | 100 mA Input Socket | |
COM | CAT |
|
|
| ||
|
|
|
| |||
|
|
|
|
|
| |
|
|
|
|
| 100 |
|
| FUSE | mA |
| |||
| F1 |
| ||||
|
| 500 |
|
| ||
| FUSED | F |
| mA |
| |
|
| 250V |
|
Fuse Holder
To remove, push in and turn counter clockwise.
To insert, reverse this procedure.
qb13f.eps
Figure
Replacing the 10 A Input Fuse
The 10 A input fuse is located inside the meter. To replace the fuse:
1.Remove the single
WWarning
Opening the case may expose hazardous voltages. To avoid electric shock, always disconnect the power cord and measuring inputs before opening the case.
2.Remove the bezel and slip the case back from the front of the meter. The fuse and fuse clip are visible at the front of the main printed circuit assembly (pca) near the input terminals.