45
Users Manual
Table
Command
RANGE1?
RANGE2?
RATE <speed>
RATE?
Description
Returns the range presently selected on the primary display.
Returns the range presently selected on the secondary display. If the secondary display is inactive, an Execution Error is generated.
Sets the measurement rate to <speed>. <speed> is either "S" for slow (2.5 readings/second), "M" for medium (5 readings/second), or "F" for fast (20 readings/second). "S", "M", and "F" can be sent as either upper- or
Returns <speed> as "S" for slow (2.5 readings/second), "M" for medium (5.0 readings/second), or "F" for fast (20 readings/second).
Measurement Queries
The commands in Table
Command
MEAS1?
MEAS2?
MEAS?
VAL1?
VAL2?
VAL?
Table
Description
Meter returns the value shown on the primary display after the next triggered measurement is complete
Meter returns the value shown on the secondary display after the next triggered measurement is complete If the secondary display is off, an Execution Error is generated.
If both displays are on, meter returns the value shown on both displays after the next triggered measurement is complete in the format selected. These values are separated by a comma (format 1); or a space, measurement units, a comma and space (format 2). See FORMAT command in Table
Example of Format 1: +1.2345E+0,+6.7890E+3<CR><LF>
Example of Format 2: +1.2345E+0 VDC, +6.7890E+3 ADC<CR><LF>
If the secondary display is not on, MEAS? is equivalent to MEAS1?
Note
If MEAS is used in external trigger (TRIGGER 2 through
TRIGGER 5), unexpected results will be obtained.
Meter returns the value shown on the primary display. If the primary display is blank, the next triggered measurement is returned.
Meter returns the value shown on the secondary display. If the secondary display is blank, the next triggered measurement is returned. If the secondary display is off, an execution error is generated.
If both displays are on, meter returns the value shown on both displays in the format selected. These values are separated by a comma (format 1); or a space, measurement units, a comma and space (format 2). See FORMAT command in Table
Example of Format 1: +1.2345E+0,+6.7890E+3<CR><LF> Example of Format 2: +1.2345E+0 VDC, +6.7890E+3 ADC<CR><LF> If the secondary display is not on, VAL is equivalent to VAL1. If a display is blank, the next triggered measurement on that display (or displays) is returned.