Operating the Meter From the Front Panel | 3 |
Selecting a Measurement Function |
Table
FUNCTION
{and FREQ
mADand FREQ
\and FREQ
e
G
All Functions
INPUT TERMINALS
VeG and COM
VeG and COM
100 mA and COM
10A and COM
VeG and COM
VeG and COM
Any terminal to earth
MAXIMUM INPUT
1000V dc
750V ac rms, 1000V peak, 2 x 107
300 mA dc or ac rms
10A dc or ac rms (or 20A dc or ac rms for 30 sec. Max)
500V dc or ac rms on all ranges
500V dc or ac rms
1000V dc or peak ac
Selecting a Measurement Function
Press a function button (white), shown in Figure
When you select a function, annunciators turn on to indicate the function selected. If a reading is shown on the secondary display when a function button is pressed, the secondary display will be turned off.
Ranges and full scale values are summarized in Table
Ranging
(E, U, T)
Ranging operations are performed using the E, U, and Tbuttons (see Figure
Autoranging
When you are in the autorange mode, the AUTO annunciator is lit.
In autorange, the meter automatically selects the next higher range when a reading is greater than full scale. If no higher range is available, "OL" (overload) is displayed on the primary or secondary display. The meter automatically selects a lower range when a read- ing is less than approximately 9 % of full scale.