MDS-JE770
Iop DATA RECORDING AND DISPLAY WHEN OPTICAL
The Iop value labeled on the optical
Record Precedure:
1.While pressing the lAMS L knob and x button, connect the power plug to the outlet, and release the lAMS L knob and x button.
2.Rotate the lAMS L knob to display “[Service]”, and press the YES button.
3.Rotate the lAMS L knob to display “Iop Write” (C05), and press the YES button.
4.The display becomes “Ref=@@@.@” (@ is an arbitrary number) and the numbers which can be changed will blink.
5.Input the Iop value written on the optical
To select the number : Rotate the lAMS L knob.
To select the digit : Press the lAMS L knob.
6.When the YES button is pressed, the display becomes “Measu=@@@.@” (@ is an arbitrary number).
7.As the adjustment results are recorded for the 6 value. Leave it as it is and press the YES button.
8.“Complete!” will be displayed momentarily. The value will be recorded in the
9.Press the REPEAT button to complete.
Display Precedure:
1.While pressing the lAMS L knob and x button, connect the power plug to the outlet, and release the lAMS L knob and x button.
2.Rotate the lAMS L knob to display “[Service]”, and press the YES button.
3.Rotate the lAMS L knob to display “Iop Read” (C26).
4.“@@.@/##.#” is displayed and the recorded contents are displayed.
@@.@ : indicates the Iop value labeled on the optical
##.# : indicates the Iop value after adjustment
5.To end, press the lAMS L button or MENU/NO button to display “Iop Read”(C26). Then press the REPEAT button.
7