Fluke 271 Set the amplitude to +10 dBm, Have a look at some other waveforms, Change the dc offset

Page 23
Set the amplitude to +10 dBm

Getting Started

Using the Instrument

Set the amplitude to +10 dBm

You can do this by keying AMPL, 1, 0, dBm.

The display will now look like this:

PER

=80.00000us

50

OUT

=+10.0dBm

DC=+0.00mV (+0.00mV)

SYM=50.0% (50.0%)

If you switch the amplitude display back to VhiZ you should see:

PER =80.00000us

VhiZ=+4.00 Vpp 50

DC=+0.00mV (+0.00mV)

SYM=50.0% (50.0%)

If you are using a 50 Ω termination on the oscilloscope input then the signal amplitude will be half the displayed value, 2 V p-p.

Have a look at some other waveforms

Using the function keys you can select any of the waveforms and observe them on the oscilloscope display.

Change the dc offset

If the output is still set to 4 V p-p then with the dc offset in the fractional millivolts range you will not be able to see changes on the oscilloscope, so the first thing to do is to force a range change.

Do this by pressing the DC OFFSET key and entering (say) 1.5 V.

You can then use the rotary control to adjust the waveform offset in steps which depend on the cursor position. For smoother control press the right DIGIT key.

PER =80.00000us

VhiZ=+4.00 Vpp 50

DC=+1.50 V (+1.50 V)

SYM=50.0% (50.0%)

Note that with the cursor placed as shown, pressing the +/- key will change the polarity of the offset voltage.

13

Image 23
Contents Programmable 10 MHz DDS Function Generator Getting Started ManualTo register your product online, visit register.fluke.com LIMITED WARRANTY AND LIMITATION OF LIABILITYSafety Getting Started Manual Alternating current Warning - hazardous voltages may be present Safety continuedThe following symbols are used on the instrument and in this manual Terminal connected to chassis ground Mains supply OFF Mains supply ONPage Immunity EMC ComplianceEmissions e EN61000-4-5 1995 Surge 0.5 kV line to line, 1 kV line to ground Title Table of ContentsAUX OUT viiiMains Supply Voltage Getting StartedIntroduction Before You StartExternally Applied Voltages shx0001f.gif Controls and ConnectionsGetting Started Figure 1. Model 271 Front PanelFunction Keys The Liquid Crystal DisplayControls FIELD and DIGIT Keys and the Rotary Control Numeric, Units and SET Keys RECALL and REMOTE MODE KeysUTILITIES Keys OUTPUT KeyAUX OUT Front Panel Inputs and OutputsEXT TRIG MAIN OUTSYNC OUT Rear Panel ConnectorsCLOCK IN/OUT VCA INTRIG/SWEEP OUT RS232 and GPIB Interface ConnectorsCheck the LCD Display Using the InstrumentPlug in and switch on the instrument Starting upSYM=50.0% 50.0% Generating Continuous SignalsSet a frequency, for example 12.5 kHz FREQ=12.50000kHz VhiZ=+20.0 Vpp 50Ω DC=+0.00mV +0.00mV=80.00000us Set the amplitude to +10 dBmHave a look at some other waveforms Change the dc offsetPER =80.00000us VhiZ=+4.00 Vpp 50Ω DC=+1.50 V +151.mV SYM=50.0% 50.0% Effect of the output attenuatorDC OFFSET CHANGE * * BY OUTPUT LEVEL Asymmetric waveforms Adjusting the phaseSOURCE=EXT TGEN=1.00ms 1.000kHz BURST COUNT= PHASE=+000 +000 SOURCE=EXT TGEN=1.00ms 1.000kHz BURST COUNT= PHASE=+116 +116TRIG/SWEEP OUT signal Generating Swept SignalsBEG FRQ=261.6000 Hz END FRQ=1.108700kHz MARK FRQ=523.3000 Hz MORE MODE=BEG-END LAW=LOG RAMP TIME=0.05 s TRIG SRC=CONTINUOUS MOREMODE=BEG-END LAW=LOG RAMP TIME=5.00 s TRIG SRC=CONTINUOUS MORE FREQ=2.000000kHz VhiZ=+20.0 Vpp 50Ω DC=+0.00mV +0.00mV SOURCE=EXT TGEN=1.00ms 1.000kHz BURST COUNT=0001 PHASE=+000 +000PHASE=+000 +000 Generating a Triggered BurstFREQ A=800.0000 Hz FREQ B=1.200000kHz SOURCE=TGEN FREE FSK ModeSpecial Waveforms StaircasesFigure 2. 625-line PAL TV Signal amplitude stepname start µsVALS=ABS AUTO=YES STEP=08 ACTIVE LENGTH=0005 LEVEL=+153 VALS=ABS AUTO=YES STEP=01 ACTIVE LENGTH=0256 LEVEL=+000RECALL ARB No ARB 03 IS EMPTY Saving and Recalling SettingsArbitrary Waveforms RECALL ARB No SINX/X ENTER TO EXECUTEOther Functions and Waveforms