Agilent Technologies E8257D/67D manual Limits

Page 28

Operation Verification

Checking the Maximum Leveled Power

Table 3-1

Frequency and Power Level Limits

 

 

 

 

 

 

 

 

 

 

 

Instrument Option

 

 

Amplitude

Start

Stop

# Points

 

 

 

Standard

Option 1EA

Frequency

Frequency

 

 

 

 

 

 

 

 

 

 

 

E8257D

 

11 dBm

 

15 dBm

250 kHz

3.2 GHz

50

 

Option 520

 

 

 

 

 

 

 

 

 

11 dBm

 

18 dBm

3.21 kHz

20 GHz

200

With Option 1E1

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

E8257D

 

7 dBm

 

14 dBm

250 kHz

3.2 GHz

50

 

Option 540

 

 

 

 

 

 

 

 

 

7 dBm

 

16 dBm

3.21 kHz

20 GHz

200

With Option 1E1

 

 

 

 

 

 

 

 

 

 

 

 

7 dBm

 

12 dBm

20.1 GHz

40 GHz

200

 

 

 

 

 

 

 

 

 

 

E8257D

 

3 dBm

 

13 dBm

250 kHz

3.2 GHz

50

Option 550 and

 

 

 

 

 

 

 

 

3 dBm

 

14dBm

3.21 kHz

20 GHz

200

 

Option 567

 

 

 

 

 

 

 

 

 

 

With Option 1E1

 

3 dBm

 

10 dBm

20.1 GHz

60 GHz

200

 

 

 

 

 

 

 

 

 

 

 

 

3 dBm

 

8 dBm

60.1 GHz

67 GHz

100

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

NOTE

For Option 1EH and Option UNW, use the data sheet to determine low band power level

 

limits.

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

24

Chapter 3

Image 28 Contents
Installation Guide Questions or Comments about our Documentation? Contents Regulatory Information Safety Information Instrument MarkingsGeneral Safety Considerations Safety InformationGetting Started Checking the Shipment Meeting Electrical and Environmental Requirements EnvironmentVentilation ChapterLine Settings Connecting the AC Power CordAC Power Cord Localization Configuring the Display Contrast and BrightnessInverse Video Screen SaverUpdating the Display During Remote Operation Gpib Interface Configuration Configuring for Remote ControlLAN Interface Configuration Dhcp Configuration Set LAN Config Manual Dhcp to DhcpRS-232 Interface Configuration Ordering Accessories Front Handles and Rack Mount FlangesE8257D/67D PSG Documentation Front Handle and Rack Mount Flange KitsOrdering Accessories Proper Usage and Cleaning Cleaning SuggestionsContacting Agilent Sales and Service Offices Returning a Signal Generator to Agilent Technologies Returning a Signal Generator to Agilent Technologies Operation Verification Performing a Self-Test Self-Test Failure Checking the Maximum Leveled Power Maximum Leveled Power Check FailureAmplitude Limits Checking the Output Power Recommended Equipment for the E8267D 540Recommended Agilent Option 532 544 Checking the Output Power Amplitude Limits Option E8257D OptionAmplitude Limits Option DBm 1E1 ±0.9 1E1 & 1EA ±1.0 50 GHz None ±1.2 E8267D Option Option 532Problems with Output Power Regulatory Information Statement of Compliance AssistanceCertification Compliance with German Noise RequirementsCompliance with Canadian EMC Requirements Compliance with Canadian EMC Requirements Index IndexIndex