Agilent Technologies E8257D/67D manual Amplitude

Page 27

Operation Verification

Checking the Maximum Leveled Power

3. Ensure that the RF OUTPUT connector is connected to a 50load.

4.Ensure that the power level entered corresponds to the value listed in Table 3-1.

5.Refer to Table 3-2for the recommended equipment and measure the output power.

If the measured power level is more than the power level listed in Table 3-1, turn the front panel knob counterclockwise until the measured power level equals the power level in Table 3-1.

6.Note the signal generator front-panel amplitude value and use it to repeat the maximum leveled power check.

If all of the above steps are followed and the check continues to fail, contact the nearest Agilent Technologies call center for assistance. Refer to “Contacting Agilent Sales and Service Offices” on page 16 for contact information.

Table 3-1

Frequency and Power Level Limits

 

 

 

 

 

 

 

 

 

 

 

Instrument Option

 

 

Amplitude

Start

Stop

# Points

 

 

Standard

Option 1EA

Frequency

Frequency

 

 

 

 

 

 

 

 

 

E8257D

 

13 dBm

 

16 dBm

250 kHz

3.2 GHz

50

Option 520

 

 

 

 

 

 

 

 

13 dBm

 

20 dBm

3.21 GHz

20 GHz

200

 

 

 

 

 

 

 

 

 

 

 

E8257D

 

9 dBm

 

15 dBm

250 kHz

3.2 GHz

50

Option 540

 

 

 

 

 

 

 

 

9 dBm

 

18 dBm

3.21 GHz

20 GHz

200

 

 

 

 

 

 

 

 

 

 

 

 

 

9 dBm

 

14 dBm

20.1 GHz

40 GHz

200

 

 

 

 

 

 

 

 

E8257D

 

5 dBm

 

14 dBm

250 kHz

3.2 GHz

50

Option 550 and

 

 

 

 

 

 

 

 

5 dBm

 

16 dBm

3.21 GHz

20 GHz

200

Option 567

 

 

 

 

 

 

 

 

 

 

 

5 dBm

 

14 dBm

20.1 kHz

60 GHz

200

 

 

 

 

 

 

 

 

 

 

5 dBm

 

12 dBm

60.1 GHz

67 GHz

100

 

 

 

 

 

 

 

 

E8267D

 

13 dBm

 

N/A

250 kHz

3.2 GHz

50

Option 520

 

 

 

 

 

 

 

 

18 dBm

 

N/A

3.21 GHz

20 GHz

200

 

 

 

 

 

 

 

 

 

 

 

E8267D

 

12 dBm

 

N/A

250 kHz

3.2 GHz

50

Option 532 and

 

 

 

 

 

 

 

 

14 dBm

 

N/A

3.21 kHz

20 GHz

200

Option 544

 

 

 

 

 

 

 

 

 

 

 

12 dBm

 

N/A

20.1 GHz

40 GHz

200

 

 

 

 

 

 

 

 

 

 

10 dBm

 

N/A

40.1 GHz

44 GHz

50

 

 

 

 

 

 

 

 

Chapter 3

23

Image 27 Contents
Installation Guide Questions or Comments about our Documentation? Contents Regulatory Information Instrument Markings Safety InformationSafety Information General Safety ConsiderationsGetting Started Checking the Shipment Chapter Meeting Electrical and Environmental RequirementsEnvironment VentilationConnecting the AC Power Cord Line SettingsAC Power Cord Localization Screen Saver Configuring the DisplayContrast and Brightness Inverse VideoUpdating the Display During Remote Operation Configuring for Remote Control Gpib Interface ConfigurationLAN Interface Configuration Set LAN Config Manual Dhcp to Dhcp Dhcp ConfigurationRS-232 Interface Configuration Front Handle and Rack Mount Flange Kits Ordering AccessoriesFront Handles and Rack Mount Flanges E8257D/67D PSG DocumentationOrdering Accessories Cleaning Suggestions Proper Usage and CleaningContacting Agilent Sales and Service Offices Returning a Signal Generator to Agilent Technologies Returning a Signal Generator to Agilent Technologies Operation Verification Performing a Self-Test Self-Test Failure Maximum Leveled Power Check Failure Checking the Maximum Leveled PowerAmplitude Limits Checking the Output Power 540 Recommended Equipment for the E8267DRecommended Agilent Option 532 544 Checking the Output Power E8257D Option Amplitude Limits OptionAmplitude Limits Option DBm 1E1 ±0.9 1E1 & 1EA ±1.0 50 GHz None ±1.2 Option 532 E8267D OptionProblems with Output Power Regulatory Information Compliance with German Noise Requirements Statement of ComplianceAssistance CertificationCompliance with Canadian EMC Requirements Compliance with Canadian EMC Requirements Index IndexIndex